Industrial Ozone Monitor - 0–10,000 ppm: Model 106-MH | 2.B Tech
The 2B Technologies Model 106-MH is an ozone monitor for concentrations from 0 to 10,000 ppm. It is designed for industrial and research applications, including analysis of water treatment off-gases before ozone destruction.
The 2B Technologies Model 106-MH is an ozone monitor for concentrations from 0 to 10,000 ppm. It is designed for industrial and research applications, including analysis of water treatment off-gases before ozone destruction.
The Model 106-MH Ozone Monitor has many uses in industry and research, such as water treatment, food safety processes and ozone disinfection. The application note available for download describes its use in a study on the effect of ozone treatments on the safety and quality of grape tomatoes. The 106-MH model was used to measure the ozone concentration in the treatment chambers. In addition, the two relays in the 106-MH were used to accurately control the ozone output of an ozone generator, ensuring that the treatments applied were well defined.
Main features
- Measurement based on UV absorption
- Low power consumption; can be battery operated (optional external lithium-ion battery)
- Internal data logger with real-time clock
- 2-second measurement interval
- On-board microprocessor with interactive menus includes data averaging options of 10 s, 1 m, 5 m, 1 hr, and user-defined adaptive filter
- USB and RS-232 output of time/date, O3 concentration, internal temperature and pressure
- Analog output (0-2.5 V and 4-20 mA) of ozone concentration in user-selected units and scaling factors
- Two 2-level relays for control purposes (e.g., control of ozone source or turn warning light on and off); Four relays on Industrial models
- Long-life pump (15,000 hr)
- Exhaust port
- Bluetooth option for wireless data transmission
- Optional configurations for 3-channel or 6-channel air sampling for Models 106-L, -M, and -MH
Technical specifications
| Measurement Principle | UV Absorption at 254 nm, single beam |
| Measurement Interval | 2 s |
| Outputs ports | USB, RS232, LCD Display, 0-2.5 V Analog, 4-20 mA |
This product the 106 series of ozone monitors was designed specifically for the industrial ozone industry to cover four different ozone concentration ranges by varying the optical path length. The ranges are: -L (low, 0-100 ppm); -M (medium, 0-1,000 ppm); -MH (medium high, 0-10,000 ppm, 0-1 vol%) and -H (0- 20 wt%, 0-14 vol%).
What should you check before choosing this product?
Match the ozone concentration range and sampling arrangement to the application. Check sample flow, measurement interval and output interfaces, and confirm the exact configuration of the instrument being supplied.