Ozone analyzer - compact low concentration monitoring: Model 108-L | 2.B Tech
Compact ozone analyzer 0–20 ppm with UV absorption, ultra-low power consumption and FEM certification for ambient, industrial and embedded monitoring applications.
The Model 108-L Ozone Monitor is a compact and energy-efficient analyzer designed for accurate ozone measurements from ppb levels up to 20 ppm. Based on UV absorption at 254 nm, it ensures reliable and precise monitoring across a wide dynamic range.
Optimized for integration, this model is specifically engineered to be embedded into existing ozone systems, using the system’s pump for sampling. Its lightweight design and very low power consumption (~2 W) make it ideal for portable, remote, or continuous monitoring applications.
The instrument is EPA approved (FEM), making it suitable for regulatory air quality monitoring and compliance applications.
Advantages
- Ultra-low detection capability: measurement from ppb to 20 ppm
- Very low power consumption: approx. 2 W for remote deployments
- Compact and lightweight: easy integration into OEM systems
- Regulatory compliance: EPA FEM approved for ambient monitoring
- Fast response: 2-second measurement interval
- Flexible outputs: analog and digital communication
- Low maintenance: robust UV absorption technology
Key Features
- Measurement principle: UV absorption at 254 nm (single beam)
- Measurement range: 0 – 20 ppm (ppb resolution)
- Resolution: 0.1 ppb
- Detection limit: 3 ppb
- Accuracy: ±2% of reading or better
- Response time: 4 seconds
- Measurement frequency: 2 s (0.5 Hz)
- Flow rate: 0.6 to 1.5 L/min
- Operating temperature: 0 to 50°C
- Power supply: 11–28 VDC (~2 W)
Typical Applications
- Ambient air monitoring: regulatory and environmental compliance
- Workplace safety: ozone exposure monitoring
- Industrial processes: ozone control and monitoring
- Ozone disinfection systems: integration and validation
- Remote monitoring stations: low-power long-term deployment
Additional Information
The Model 108-L provides RS232 and analog outputs (0–2.5 V, 4–20 mA) with user-defined scaling, enabling seamless integration into industrial control systems.
Its relay with configurable setpoints allows automated process control or alarm triggering. The analyzer also includes temperature and pressure compensation and configurable data averaging options.
Optional configurations include enclosure, integrated pump, and AQLite outdoor packages, making the system adaptable for both fixed and mobile monitoring applications.